3 determining samples, Determining samples – Metrohm 860 KF Thermoprep User Manual

Page 42

Advertising
background image

5.3 Determining samples

■■■■■■■■■■■■■■■■■■■■■■

34

■■■■■■■■

860 KF Thermoprep

Note

Caution! The vessel is hot.

4 Place the blank into the oven using the crimping tongs.

5 Use both hands to move the guide head downward and insert the

needle into the blank.

Waiting for the determination

1 Wait until the determination is finished. Conditioning will then be

resumed automatically.

2 Repeat the blank value determination with the next blank.

5.3

Determining samples

The blank value must be subtracted from the calculated consumption
when determining a sample.

Preparing the KF titrator

1 Load a suitable method.

2 Start the conditioning by pressing the [Start] key.

Preparing the sample

1 Weigh the sample into a sample vessel and seal it tightly with the

crimping tongs.

2 Check that the oven temperature and the gas flow rate correspond

to the values which have been set.

3 Wait until the working medium in the KF titration vessel has been

conditioned and the signal drift is stable.