4.5 Parameters

■■■■■■■■■■■■■■■■■■■■■■

26

■■■■■■■■

860 KF Thermoprep

4.5

Parameters

[OK]

The instrument parameters can be adjusted in a menu that is opened by
pressing the [OK] key. Exit the Parameters menu with [QUIT].

Note

Modified values are applied after confirmation with the [OK] key. A
modification is discarded with [QUIT].

Temp. 1

Set value for temperature 1.

Range

50 ... 250 °C (Increment: 1)

Default value

100 °C

Temp. 2

Set value for temperature 2.

Range

50 ... 250 °C (Increment: 1)

Default value

100 °C

Gas flow

Set value for the gas flow.

Range

10 ... 150 mL/min (Increment: 1)

Default value

50 mL/min

Gas type

Selection of the gas.

Selection

air | nitrogen | other

Default value

air

other
If other is selected, then a measurement factor for the gas that is used
must be specified. The measuring factor is used for the correct meas-
urement of the flow rate.

Meas. factor

The measuring factor for measuring the flow rate of the gas. This setting
is only possible if other is selected for the gas. The corresponding measur-
ing factor is automatically applied for air and nitrogen.