3.5 Capillary connections in the IC system

■■■■■■■■■■■■■■■■■■■■■■

16

■■■■■■■■

882 Compact IC plus – Anion

3.4.3

Location

The instrument has been developed for operation indoors and may not be
used in explosive environments.

Place the instrument in a location of the laboratory which is suitable for
operation, free of vibrations, protected from corrosive atmosphere, and
contamination by chemicals.

The instrument should be protected against excessive temperature fluctu-
ations and direct sunlight.

3.5

Capillary connections in the IC system

This chapter contains general information concerning the capillary con-
nections in the IC instruments and systems.

Generally speaking, capillary connections between two components of an
IC system are made up of one connection capillary and two pressure
screws with which the capillary is connected to the respective compo-
nents.

Pressure screws

4

1

2

3

Figure 4

Connection of capillaries with pressure screws

1

PEEK pressure screw (6.2744.014)
Use on the injection valve.

2

Connection capillary

3

PEEK pressure screw, short
(6.2744.070)
For use on the high pressure pump, the
purge valve, the inline filter, the pulsation
absorber, the guard column and the separa-
tion column.

4

PEEK pressure screw, long
(6.2744.090)
Use on special components. Is not used on
all instruments.