1 connecting the suppressor, 14 metrohm suppressor module (msm), Connecting the suppressor – Metrohm 882 Compact IC plus – Anion User Manual

Page 52: Figure 24, Suppressor – connection capillaries

Advertising
background image

3.14 Metrohm Suppressor Module (MSM)

■■■■■■■■■■■■■■■■■■■■■■

42

■■■■■■■■

882 Compact IC plus – Anion

3.14

Metrohm Suppressor Module (MSM)

The MSM is used for chemical suppression during anion analysis with con-
ductivity detection or UV/VIS detection. It consists of 3 suppressor units in
total, which are, in rotation, used for suppression – regenerated with
100 mmol/L sulfuric acid – rinsed with ultrapure water.

Suppression reaction in the MSM
When using a carbonate eluent, the following reaction (amongst others)
occurs in the MSM:

2R-SO

3

-

H

+

+ Na

2

CO

3

➙ 2R-SO

3

-

Na

+

+ H

2

O + CO

2

3.14.1

Connecting the suppressor

The three inputs and outputs of the suppressor units, numbered with 1, 2
and 3 on the connecting piece, each have 2 fixed mounted PTFE capilla-
ries.

1

3

2

3

4

5

6

7

8

2

1

Figure 24

Suppressor – connection capillaries