3 cleaning the suppressor – Metrohm 882 Compact IC plus – Anion User Manual

Page 86

Advertising
background image

5.11 Metrohm Suppressor Module (MSM)

■■■■■■■■■■■■■■■■■■■■■■

76

■■■■■■■■

882 Compact IC plus – Anion

Connect the inlet capillary for the rinsing solution (labeled with
rinsing solution) to the outlet of the high pressure pump with the
aid of a coupling (6.2744.040).

Rinse the first suppression unit with degassed ultra pure water for
about 15 minutes.

In the software, use the Step command to switch to the second
suppressor unit and rinse this for about 15 minutes.

In the software, use the Step command to switch to the third
suppressor unit and rinse this for about 15 minutes.

5 Connect the suppressor to the IC system

Reconnect the capillaries labeled regenerant and rinsing solution
to the IC system.

5.11.3.3

Cleaning the suppressor

In the following cases, it may be necessary to clean the suppressor:

Increased backpressure onto the connection tubings of the suppres-
sor..

Blockage of the suppressor which cannot be eliminated (solutions can
no longer be pumped through the suppressor).

Jamming of the suppressor which cannot be eliminated (suppressor
can no longer be switched over).

Cleaning the suppressor

Clean the suppressor as follows:

1 Disconnecting the suppressor from the IC system

Switch off the instrument.

Disconnect all capillaries of the suppressor from the IC system.

2 Dismantling the suppressor

Unscrew union nut (36-1) from the housing (36-4).

Pull the connecting piece (36-2) and the rotor (36-3) out of the
housing.
The connecting piece and the rotor normally stick to one another
– if this is not the case: Take a sharp object, insert it into the slot
of the housing (36-5) and pull out the rotor in this way.

Detach the connecting piece from the rotor.