Selecting the sample loop
The amount of sample solution injected depends on the volume of the
sample loop. The choice is made on the basis of the application. The fol-
lowing sample loops are normally used:
Cation determination
10 µL
Anion determination with suppression
20 µL
Anion determination without suppression
100 µL
3.13
Peristaltic pump
3.13.1
Principle of the peristaltic pump
The Peristaltic pump is used for pumping sample and auxiliary solutions. It
can rotate in both directions.
The peristaltic pump pumps liquids according to the principle of displace-
ment. The pump tubing is clamped between the rollers (20-3) and the
tubing cartridge (20-5). During operation, the peristaltic pump drive
rotates the roller hub (20-2), so that the rollers (20-3) push the liquid for-
ward in the pump tubing.