3 selecting the sample loop, 13 peristaltic pump, 1 principle of the peristaltic pump – Metrohm 882 Compact IC plus – Anion User Manual

Page 47: Selecting the sample loop, Principle of the peristaltic pump, Figure 20, Peristaltic pump

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

882 Compact IC plus – Anion

■■■■■■■■

37

3.12.3

Selecting the sample loop

The amount of sample solution injected depends on the volume of the
sample loop. The choice is made on the basis of the application. The fol-
lowing sample loops are normally used:

Cation determination

10 µL

Anion determination with suppression

20 µL

Anion determination without suppression

100 µL

3.13

Peristaltic pump

3.13.1

Principle of the peristaltic pump

The Peristaltic pump is used for pumping sample and auxiliary solutions. It
can rotate in both directions.

The peristaltic pump pumps liquids according to the principle of displace-
ment. The pump tubing is clamped between the rollers (20-3) and the
tubing cartridge (20-5). During operation, the peristaltic pump drive
rotates the roller hub (20-2), so that the rollers (20-3) push the liquid for-
ward in the pump tubing.

6

1

4

3

2

5

7

Figure 20

Peristaltic pump

Advertising