5 Operation and maintenance

772 Pump Unit

■■■■■■■■

17

5.2.2

Suitability test for tubing materials

Place a piece of the tubing in question for 48 hours in a closed container
filled with the liquid in question. Then examine the piece of tubing for
signs of swelling, embrittlement or other damage.

5.2.3

Tubing dimensions

Note

Use tubing with a wall thickness of 1.6 mm only.

The largest usable inner diameter is 6.4 mm.

5.2.4

Tubing lifetime

The pump tubing has a limited lifetime and therefore has to be replaced
regularly. The following are examples of factors that can influence the life-
time of tubing in peristaltic pumps:

Chemical resistance of the pump tubing

Pump rate (rotational speed)

Properties of the pumped liquid

Outlet pressure

Ambient temperature

Shearing force of the rollers

Twisted or kinked tubing

Tubing too long

Wall thickness of 1.6 mm not complied with