5.2 Pump tubings

■■■■■■■■■■■■■■■■■■■■■■

18

■■■■■■■■

772 Pump Unit

Note

If the tubing is not tightened sufficiently around the rotor, this has a
negative impact on the lifetime of the tubing, as in this case the tub-
ing tends to fold up. If the tubing is tightened too much, the flow is
restricted.

The tubing will last longer if the pressure clamp is detached when
the pump is not in use.

In order to achieve good repeatability of flow rate following a
change in tubing, it is imperative that you always select the same
tubing length.

Before changing tubing:

Pump all of the liquid out of the tubing so that none remains.

Pull out the connection cable of the peristaltic pump to prevent acci-
dental starting of the pump.

Detach the pressure clamp, open the tubing clamps and remove the
tubing.

5.2.5

PharMed® pump tubing data sheet

Physical properties

Polypropylene-based thermoplastic elastomer, contains
plasticizer

Dimensions

Inner diameter = 6.4 mm

Wall thickness 1.6 mm

Advantages

Ideal for aqueous media

Impermeable to normal light and UV radiation

Tubing can be welded, glued and thermoformed.

Very long lifetime

Restrictions

Additives can be dissolved out by the pumped medium.

Applications:

Acids

Bases

Solvents

Pressure

Vacuum

Viscous media

Sterile media

Good resistance

Good resistance

Not suitable

Well suited

Very well suited

Well suited

Very well suited

Temperature range

–60 - +130 °C