6 viton® pump tubing data sheet, Viton® pump tubing data sheet, Figure 10 – Metrohm 772 Pump Unit User Manual

Page 27: Calculation formula for the permeability (c)

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

772 Pump Unit

■■■■■■■■

19

Permeability (C) for:

(see Figure 10, page 19)

CO

2

O

2

N

2

1,200

200

80

Corresponds to the fol-
lowing standards:

USP Class VI

FDA 21CFR177.2600

NSF (Standard 51)

Cleaning/sterilization

The tubing can be sterilized in the autoclave without
being damaged.

Figure 10

Calculation formula for the permeability (C)

5.2.6

Viton® pump tubing data sheet

Physical properties

Viton®; rigid, impermeable, black

Dimensions

Inner diameter = 6.4 mm

Wall thickness 1.6 mm

Advantages

High chemical resistance

Low gas permeability

Wide temperature range

Restrictions

Limited lifetime

Applications:

Acids

Bases

Solvents

Pressure

Vacuum

Viscous media

Sterile media

Outstanding

Outstanding

Limited

Not recommended

Good

Good

Satisfactory

Temperature range

–31 °C - +204 °C

Permeability (C) for:

CO

2

O

2

N

2

38

14

5

Corresponds to the fol-
lowing standards:

None

Cleaning/sterilization

Not recommended