5 Troubleshooting

850 Professional IC – AnCat

■■■■■■■■

97

Problem

Cause

Remedy

MSM – regeneration or
rinsing solution flow prob-
lems.

Check the flow of the regeneration and rinsing
solution (see Chapter 2.16.2, page 47).

Data of the separa-
tion column cannot
be read.

Column chip contamina-
ted.

Clean contact surfaces of the column chip
(with alcohol).

Column chip defective.

1. Save column configuration in MagIC Net.
2. Notify Metrohm Service.

Individual peaks
greater than expec-
ted

Sample – Carry-over of the
samples from previous
measurement.

Rinse system longer between two samples.

MSM – No (or insuf-
ficient) pumping of
regeneration or rins-
ing solution

Leak in the system.

Check connections.

Peristaltic pump – contact
pressure too weak.

Correctly set contact pressure (see "Set flow
rate", page 54)
.

Peristaltic pump – filter
blocked (see Figure 29,
page 52).

Replace the filter (see "Replacing the filter",
page 91)
.

MSM – backpressure too
high.

Clean the MSM (see Chapter 4.11.3.2, page
85)
or replace parts
(see Chapter 4.11.3.3,
page 88)
.

Peristaltic pump – pump
tubing defective.

Replace pump tubing (see Figure 28, page 51).

Poor reproducibility
of the retention
times

Eluent – Leak in the eluent
path.

Check the eluent path.

Eluent – Blockage in the
eluent path.

Check the eluent path.

Chromatograms
have poor resolution

Separation column –
diminished separating effi-
ciency.

Regenerate separation column (see Chap-
ter 4.14.4, page 93)
.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 61)
.

Conductivity detec-
tor is not recognized
in the software

No connection.

Check the cable connection (32-

1

).

Turn the instrument off and on again (after
15 seconds).