Figure 32, Conductivity detector rear – Metrohm 850 Professional IC AnCat User Manual

Page 66

Advertising
background image

2.18 Conductivity detector

■■■■■■■■■■■■■■■■■■■■■■

56

■■■■■■■■

850 Professional IC – AnCat

1

3

2

Figure 32

Conductivity detector rear

1

Detector cable
With mounted plug.

2

Detector output capillary
Permanently installed.

3

Type plate
With serial number.

Note

In order to prevent unnecessary peak widening after separation, the
connection between the outlet of the separation column and the inlet
in the detector should be kept as short as possible.

Connecting the detector input capillary to the MSM

1

Connect the detector input capillary (33-1) and the capillary of
the MSM (33-2) marked with Detector to one another using one
6.2744.040 coupling (33-3) and two 6.2744.070 short pressure
screws (33-4).