13 conductivity measuring system – Metrohm 850 Professional IC AnCat User Manual

Page 114

Advertising
background image

6.13 Conductivity measuring system

■■■■■■■■■■■■■■■■■■■■■■

104

■■■■■■■■

850 Professional IC – AnCat

6.13

Conductivity measuring system

Type

Microprocessor-controlled Digital Signal Processing (DSP technol-
ogy)

Intelligent detector with 6 reference chromatograms

Measuring range

0…15000 µS/cm single range

Noise

< 0.1 nS at 1 µS/cm

Deviations from
the linearity

< 0.1 % for conductivity values greater than 16 µS/cm (typical for
analyses with MSM)

Drift

< 0.2 nS/cm per hour

Measuring rate

10 measurements per second for optimum results without filtering

Resolution

0.0047 nS/cm

Baseline

Noise < 0.2 nS/cm typical for chemical suppression

Conductivity
detector

Cell volume

0.8 µL

Cell constant

Individual calibration data saved in the detector

adjustable range: 13.0…21.0 /cm

Electrodes

Ring-shaped electrodes made from stainless steel

Materials in
contact with
eluent

Chemically inert PCTFE

Maximum oper-
ating pressure

5.0 MPa (50 bar)

Cell tempera-
ture

20…50 °C in increments of 5 °C

Temperature
stability

< 0.001 °C

Temperature
compensation

0…5 %/K adjustable, default 2.3 %/K

Heating up time

< 30 minutes (40 °C)

Advertising