3 status and error messages – Metrohm 768 KF Oven User Manual

Page 21

Advertising
background image

3.1 General rules3.1 General rules

768 KF Oven

18

3.1.3 Status and error messages


Detailed status conditions

Status conditions of the global $G:
$G.Mode .Inac

Waiting during start delay

.PurgeTime

Waiting during purge time

.CondTime

Waiting during cond.time

.HeatSmpl

Heating the sample

.Terminate

Carrying out the terminating steps

$G.Assembly .Prep.Wait

Heating the oven to the set temperature

.Boat

The sample boat has been manually moved


Status conditions of the global $R:
$R.Mode .Ready

Inactive: Ready to start an automatic determination

$R.Assembly.Ready

An Assembly step has been carried out. Ready to carry out
another Assembly step (&Mode $G will trigger error message
E31).


Status conditions of the global $S:
The instrument gives the status from which it has been stopped. The detailed status information is
therefore identical as for the global status $G.


Error messages

RS receive errors:

E36

Parity error
Exit: <QUIT> and set parity at both devices the same.

E37

Stop Bit
Exit: <QUIT> and set stop bit at both devices the same.

E38

Overrun error. At least 1 character could not be read.
Exit: <QUIT>.

E39

The internal receive buffer is full (>82 characters).
Exit: <QUIT>.