3.2 Remote control commands3.2 Remote control commands
768 KF Oven
26
3.2.2.21
Info.ActualInfo.Meas.CyclNo
read only
Info.ActualInfo.Meas.SampleTemp
read only
Info.ActualInfo.Meas.OvenTempread only
Info.ActualInfo.Meas.GasFlowread only
Inquiry of the current values.
NV: Not Valid.
If the measured value is exceeded, OV (overrange) is sent.
A time frame can be generated from the cycle number and the cycle time (see 3.2.2.24).
The cycle number is always zeroed on start and on termination of an automatic
determination.
3.2.2.22
Info.ActualInfo.Status.BoatPosread only
Info.ActualInfo.Status.Valve read only
Info.ActualInfo.Status.Pump read only
Info.ActualInfo.Status.Heating read only
Status inquiry of the individual assemblies in the following format (examples):
.BoatPos
Position of the sample boat
23 mm
.Valve
Position of the valve
"purge" or "transfer"
.Pump
Status of the pump
"ON" or "OFF"
.Heating
Heating level
20
3.2.2.23
Info.ActualInfo.Display.L1
up to 24 ASCII characters
Info.ActualInfo.Display.L2
up to 24 ASCII characters
1st and 2nd line of the display. The display can be written to.
The display is not operated by the KF oven if &Setup.Lock.Display is set to ON, see
3.2.2.35.
$Q sends the contents of the corresponding display line.
3.2.2.24
Info.Assembly.CycleTime
read only
Inquiry of the cycle time in s.
3.2.2.25
Assembly.Prep
$G, $S
Switch heating to the sample temperature on/off.
3.2.2.26
Assembly.Heat
$G
Assembly.Heat.Value
0...50
Control heating. The set heating power level is started with &Assembly.Heat $G.
0 means switch off heating.
3.2.2.27
Assembly.Valve
$G
Assembly.Valve.Pos
purge, transfer
Set valve. The selected position is set with &Assembly.Valve $G.
3.2.2.28
Assembly.Boat
$G, $S
Assembly.Boat.Rate
0.1...10
Assembly.Boat.Pos
0...130.0
Assembly.Boat.SetPos.InPos
0...130.0
Assembly.Boat.SetPos.OutPos0...130.0
The movements of the sample boat are initiated with &Assembly.Boat $G.