3.2 Remote control commands3.2 Remote control commands

768 KF Oven

24

3.2.2.8

Config.OvenSet.CharSet

Epson, Seiko, Citizen,

HP, IBM

Selection of the character set.
IBM means IBM character set according to character set table 437. Select "IBM" for work
with the computer.


3.2.2.9

Config.OvenSet.Report

ON, OFF

ON means: Output of a report at the end of the automatic determination.


3.2.2.10

Config.Aux.Language

english, deutsch, francais, español

Selection of the dialog language.


3.2.2.11

Config.Aux.RunNo

0...9999

Current run number.
Is set to 0 on power on and on initialisation. Counting starts again at 1 after 9999 has been
reached.


3.2.2.12

Config.Aux.AutoStart

1...9999, OFF

Config.Aux.StartDelay

0...9999

.AutoStart:

Number of automatic, internal starts for automatic determinations.

.StartDelay:

Start delay time in s. During this time, the data of the preceding determination
are retained.



3.2.2.13

Config.Aux.Beeper

1...9, OFF

Config.Aux.DevName

up to 8 ASCII characters

Number of beeps if the Oven is ready and after the determinations.
Name of the device for interconnections with several devices. It is advisable to use only the
characters A...Z (ASCII No. 65...90), a...z (ASCII No. 97...122) and 0...9 (ASCII No. 48...57)
if the Setup.AutoInfo function (see 3.2.2.38) has been activated at the same time.


3.2.2.14

Config.Aux.Prog

read only

Output of the program version.
In response to $Q, the KF Oven sends: "5.768.0010".


3.2.2.15

Config.RSSet

$G

$G sets all RS settings. After setting the interface parameters, wait at least 2 s to allow the
components to equilibrate.


3.2.2.16

Config.RSSet.Baud

300, 600, 1200, 2400, 4800, 9600

Config.RSSet.DataBit

7, 8

Config.RSSet.StopBit

1, 2

Config.RSSet.Parity

even, odd, none

Config.RSSet.Handsh

HWs, HWf, SWchar, SWline, none

Setting of the RS interface parameters: Baud rate, data bits, stop bits, parity and type of
handshake, see also page 31.
Setting of the values must be initiated immediately following the entry with &Config.RSSet
$G, see 3.2.2.15.