3 Installation
944 Professional UV/VIS Detector Vario (2.944.0010)
■■■■■■■■
17
3.5
Connecting the flow-through cell
To connect the capillaries to the flow-through cell, proceed as follows:
Connecting the capillaries
1 Connecting the detector input
2
1
3
Figure 5
Connecting the detector input
1
Detector input capillary
PEEK capillary (6.1831.100).
2
Detector inlet
Labeled IN.
3
Pressure screw
■
Unscrew the pressure screw from the IN detector inlet.
■
Slide the presssure screw over the detector input capillary such
that a small part of the capillary is visible at the top.
■
Tighten the detector input capillary in the detector inlet using the
pressure screw.