Metrohm 944 Professional UV/VIS Detector Vario User Manual

Page 38

Advertising
background image

6.7 Cleaning the flow-through cell

■■■■■■■■■■■■■■■■■■■■■■

30

■■■■■■■■

944 Professional UV/VIS Detector Vario (2.944.0010)

2 Connect the input capillary to a high-pressure pump or a peristaltic

pump and rinse the flow-through cell as described below; make sure
that the permissible maximum pressure of 5 MPa is not exceeded:

first rinse with ultrapure water to avoid precipitates,

then rinse several minutes with methanol, to dissolve the deposits,
and

finally, rinse with ultrapure water for at least another 15 minutes
to wash away the dissolved deposits.

3 Observe the baseline during the last rinsing with ultrapure water.

If the baseline is only slightly noisy, then the flow-through cell is
clean.

If the noise on the baseline persists, rinse the flow-through cell once
again with another solvent (see "Rinsing the flow-through cell with
another solvent", page 30)
.

Rinsing the flow-through cell with another solvent

A solvent made up of a 1:2 mixture of acetic acid and isopropanol has
shown good results. Depending on the application, other solvents may
also prove efficient.

Prerequisites:

Rinsing with methanol was ineffective.

The input capillary is connected to a high-pressure pump or a peristaltic
pump.

1 Rinse the flow-through cell as described below; make sure that the

maximum permissible pressure of 5 MPa is not exceeded:

first rinse with ultrapure water to avoid precipitates,

then rinse several minutes with the selected solvent to dissolve
the deposits, and

finally, rinse with ultrapure water for at least another 15 minutes
to wash away the dissolved deposits.

2 Observe the baseline during the last rinsing.

If the baseline is only slightly noisy, then the flow-through cell is
clean.

If the noise on the baseline persists, the flow-through cell must be
dismantled and cleaned manually (see "Dismantling and cleaning the
flow-through cell", page 31)
.

Advertising