6 Operation and maintenance

944 Professional UV/VIS Detector Vario (2.944.0010)

■■■■■■■■

31

Dismantling and cleaning the flow-through cell

NOTE

Ex works, the retaining screws are tightened with a torque wrench. This
guarantees a pressure stability up to 5 MPa (50 bar).

Once the flow-through cell has been opened and closed again, this
maximum pressure stability can no longer be guaranteed.

Prerequisites:

Rinsing the flow-through cell with solvent was ineffective.

Required tools:

To open the measuring cell, you need a slotted screwdriver, size 5.

Seals (6.2764.000).

1 Removing the flow-through cell

Unscrew the input capillary and the output capillary.

Loosen and remove the knurled screws.

Remove the flow-through cell from the optical block.

5

4

3

1

4

3

1

2

6

Figure 12

Flow-through cell – parts

1

Retaining screw

2

Outer seal

3

Lens

4

Inner seal

5

Measuring cell – opening

6

Cell holder