3.5 Connecting the flow-through cell
■■■■■■■■■■■■■■■■■■■■■■
18
■■■■■■■■
944 Professional UV/VIS Detector Vario (2.944.0010)
2 Connecting the detector outlet
3
4
1
2
Figure 6
Connecting the detector outlet
1
Detector input capillary
2
Detector output capillary
PEEK capillary (6.1831.100).
3
Detector outlet
Labeled OUT.
4
Pressure screw
■
Unscrew the pressure screw form the OUT detector outlet.
■
Slide the presssure screw over the detector output capillary such
that a small part of the capillary is visible at the top.
■
Screw the detector output capillary to the detector outlet with the
pressure screw.
NOTE
The detector output capillary (6.1831.100) is 1 m long and may
not be shortened.