3.5 Titration vessel for coulometric KF titration

■■■■■■■■■■■■■■■■■■■■■■

30

■■■■■■■■

851 Titrando

6 Connect the tubing for the aspiration of the coulometer cell at the

lower connector of the addition and aspiration tube (6).

3.5.4

Coulometer cell with aspiration equipment (utilization with Dos-
ino)

A Dosino allows the automatic replacement of reagents.

For aspiration, the 6.5617.000 aspiration equipment is used, including a
complete dosing unit and a 50 mL glass cylinder. For aspirating greasy
samples when only the sample and not the whole reagent is aspirated, we
recommend using a dosing unit with a 20 mL cylinder. For highly viscous
samples a dosing unit with a 10 mL cylinder is suitable.

Mounting the aspiration tip

2

6.1543.070

3

1

6.1446.060

6.2730.030

Insert the aspiration tip into the coulometer cell as follows:

1 Screw the 6.2730.030 nipple, with stopper and O-ring, onto the

6.1446.060 stopper.

2 Pull out the stopper.

3 Slide the 6.1543.070 aspiration tip through the stopper.