8 technical specifications, 1 measuring interface, 1 generator electrode – Metrohm 851 Titrando User Manual

Page 60: 2 indicator electrode, 3 temperature, Measuring interface, Generator electrode, Indicator electrode, Temperature

Advertising
background image

8.1 Measuring interface

■■■■■■■■■■■■■■■■■■■■■■

52

■■■■■■■■

851 Titrando

8 Technical specifications

8.1

Measuring interface

The 851 Titrando has one galvanically isolated measuring interface.

The measuring cycle is 100 ms for all measuring modes.

8.1.1

Generator electrode

One measuring input (Gen.) for a generator electrode.

I

max

400 mA
Continuous current and pulsed

8.1.2

Indicator electrode

One measuring input (Ind.) for an indicator electrode.

Measuring mode
Ipol

Determination with adjustable polarization current

AC

5, 10, 20 and 30 µA

DC

–125 - +125 µA

8.1.3

Temperature

A measuring input (Temp.) for temperature sensors of the Pt1000 or NTC
type with automatic temperature compensation.

R (25 °C) and B value can be configured for NTC sensors.

Measuring range

Pt1000

–150 - +250 °C

NTC

–5 - +250 °C
(R (25 °C) = 30,000

Ω and B (25/50) = 4,100 K)

Resolution

Pt1000

0.1 °C

NTC

0.1 °C

Measuring accu-
racy

Pt1000

±0.2 °C
(Applies for measuring range –20 - +150 °C; ±1 digit; without sensor
error, under reference conditions)

NTC

±0.6 °C