5 ambient temperature, 6 interfaces, Ambient temperature – Metrohm 851 Titrando User Manual

Page 62: Interfaces

Advertising
background image

8.5 Ambient temperature

■■■■■■■■■■■■■■■■■■■■■■

54

■■■■■■■■

851 Titrando

NE 21

8.5

Ambient temperature

Nominal function
range

+5 - +45 °C

Storage

–20 - +60 °C

Transport

–40 - +60 °C

8.6

Interfaces

USB connectors

USB ports

2 USB downstream ports (type A sockets), 500 mA each, for connect-
ing peripheral devices such as printers, keyboards, barcode readers or
RS-232/USB boxes (Metrohm order no. 6.2148.020).

Connector "Controller"

Controller port

USB upstream port with auxiliary power supply (Mini DIN socket) for
connecting Touch Control or computer for controlling the 851
Titrando.

Touch Control

With integrated Touch Control cable.

Computer

With 6.2151.000 cable.

MSB connectors (Metrohm Serial Bus)

Dosing device

Connection for a maximum of 4 external dosing devices, models Dosi-
mat or Dosino (MSB 1 to MSB 4).

Stirrer

Connection for a maximum of 4 stirrers.
Stirrer control: switching on/off manually or coordinated with the titra-
tion sequence.
Speed in 15 steps and shift direction can be selected.

Remote Box

Connection for a maximum of four Remote Boxes. Remote Boxes can
be used to actuate and monitor external devices.

Advertising