2.6 Parameters, key <PARAM>

799 GPT Titrino

15

2.6 Parameters, key <PARAM>

The key <PARAM> is used for the entry of values that

determine the modes. Values marked with "cond." are

accessible during the conditioning in the SET and KFT

mode. "**titr." means that these values can be

changed during the titration. They influence the ongo-

ing determination. Other values can only be changed in

the inactive state.

The display texts of the Titrino are shown to the left.

The values are the default values.

2.6.1 Parameters for DET and MET









titration parameters

determine the course of the titration and measured

value acquisition.

stop conditions:

Parameters for the automatic termination of the titra-

tion.

statistics:

Calculation of mean values and standard deviation,

see page 77.

evaluation:

Parameters for the evaluation of EP's, fix EP's, and

pK/HNP.

preselections:

ON/OFF of various auxiliary functions such as auto-

matic requests after the start and activate pulse.


>titration parameters

meas.pt.density 4
DET



min.incr. 10.0 µl
DET


V step 0.10 ml
MET





General titration parameters

Measuring point density (0...9)

0 means highest density, 9 lowest. Selection of the

meas.pt.density, see page 21.

Minimum increment (0...999.9 µL)

The increment is dispensed at the beginning of the

titration and in the region of the equivalence point.

Size of volume increment (0...9.999 mL)

Dosing step. Small volume increments are needed to

determine blank values or to assure accuracy with

highly unsymmetrical curves. Selection of size of the

increment, see page 21.

PARAM

parameters

>titration parameters

>stop conditions

>statistics

>evaluation

>preselections