6 parameters for doc – Metrohm 799 GPT Titrino User Manual

Page 66

Advertising
background image

2.6. Parameters, key <PARAM>

799 GPT Titrino

62

2.6.6 Parameters for DOC









dosing parameters

control the dosing

stop conditions:

Parameters for termination.

statistics

:

Calculation of mean values and standard deviation,

see page 77.

monitoring

of limit values for the measured value and temperature.

preselections

:

ON/OFF of various auxiliary functions such as auto-

matic requests after the start and activate pulse.


>dosing parameters

begin at pH init






end at pH OFF






sweep time 300 s



dynamics 0.25
**titr.















Dosing parameters

Start of the measured value for the beginning of dosing

(input range depends on the measured quantity:

pH: 0...±20.00,

init

U, Ipol: 0...±2000 mV, init

Upol:

0...±200.0 µA, init)

<CLEAR> sets "init" = initial measured value.

Final measured value (input range depends on the

measured quantity:

pH: 0...±20.00,

OFF

U, Ipol: 0...±2000 mV, OFF

Upol:

0...±200.0 µA, OFF)

<CLEAR> sets "OFF"

Sweep time for measured value (0...999 999 s)

The change in the set measured value from the initial to

the final value within the sweep time is linear.

Control range (input range depends on the measured

quantity:

pH: 0...20.00,

OFF

U, Ipol: 0...±2000 mV, OFF

Upol:

0...±200.0 µA, OFF)

<CLEAR> sets "OFF".

"OFF" means largest control range (dynamics), i.e.

slow reagent addition.

Outside the control range, dispensing is performed

continuously, see also page 68.

parameters

>dosing parameters

>stop conditions

>statistics

>monitoring

>preselections