5.1 Setting up and connecting the instruments

799 GPT Titrino

211

inputted from the balance. For this, the address of the identifications

and method, resp. must be preselected on the input unit.

Balance

Method

Id1 Id2 Id3

Sartorius

METH or 27

ID.1 or 26

ID.2 or 24

C-20 or 23

Mettler (AT)

D (Mthd)

C (ID#1) B

(ID#2)

A (c20)

If balance and printer are connected at the same Titrino COM you

need the 6.2125.010 and 6.2125.030 Adapters.

If the balance works only with 7 bit and the printer with 8 bit and if

hey are at the same Titrino COM, the balance has to be set to

"space parity" and Titrino/printer to 8 bit, "no parity".

5.1.5 Connection of a Sample Changer

With 6.2141.030 cable (instead of 6.2141.020), two Titrinos can be

connected to the 730 or 760 Sample Changer at the same time.


• The "Remote" socket allows not only connection of a sample

changer but also additional control functions. Pin assignment of

the "Remote" socket and control possibilities, see page 223.

• If a calibration has to be performed with the sample changer, the

calibration parameter "sample changer:" must be set to "ON".

• In connections with the sample changer, "auto start" should be

set to "OFF" in the <CONFIG> key. The start command is given

by the Sample Changer.

730/760

6.2141.020 Cable

C

B

A

D