3 equipment required, 4 diagnosis steps, 1prepare instruments for diagnostic test – Metrohm 799 GPT Titrino User Manual

Page 198: 2perform display test

Advertising
background image

4.2. Diagnosis

799 GPT Titrino

194

4.2.3 Equipment required:

voltage calibrator,

e.g. 1.773.0010 Metrohm pH Simulator

or 1.767.0010 Calibrated Reference for mV, pH, Ω

µ

S, °C

highly insulated interconnection cable

6.2108.060

resistor switch-box, class 0.1 % (or resistor 14.3 k, 0.1 %)
cable 3.496.5070
exchange units, if possible with different cylinder volumes (or 3.496.0070 dummy exchange unit)
stop watch or watch with second hand
digital or analogue voltmeter (if need be, connect a calibrated recorder)
2 connecting cables with 4 mm banana plugs
test plug 3.496.8550 (necessary only if plug ‘Remote’ should be checked)
test plug 3.496.8560 (necessary only if plug ‘RS 232’ should be checked)

4.2.4 Diagnosis steps

1

Prepare instruments for diagnostic test

• Power off.

• Disconnect all external connections (cables at rear, except mains cable and key-

board).

• Remove exchange unit.

• Power on and immediately press and hold the <9> key until the powerup test pat-

tern disappears.

diagnose press key 0...9

2

Perform display test

• Press <2>.

display test

• Press <Enter>.

Characters for a visual check of the display are generated on the eight lines.

Test sequence:

a) The display is cleared and overwritten from the top left to the bottom right with a dot

pattern ( ▓▓▓▓▓▓▓▓ ).

b) The display is cleared and overwritten from the top left to the bottom right with a dot

pattern( ░░░░░░░░.

c) The display is continuously cleared and overwritten from the top left to the bottom right

with the complete character set (see page 195 / Fig. 1). At the same time with moving

display the LED’s „COND.“, „STATISTICS“ and „SILO“ are swtiches on and off.

• The test sequence can be held and then continued at any time by pressing <5>.

• Block 2 is quit by pressing <Clear>.

diagnose press key 0...9