6.1
■■■■■■■■■■■■■■■■■■■■■■
102
■■■■■■■■
930 Compact IC Flex Oven/SeS/PP/Deg (2.930.2560)
Problem
Cause
Remedy
Vacuum is not being
built
Eluent degasser – Vacuum
connection on the rear of
the instrument is not
(tightly) sealed.
■
Seal the Vacuum connector tightly with a
threaded stopper (6.1446.040).