2 connecting the mcs, Connecting the mcs, Figure 15 – Metrohm 930 Compact IC Flex Oven/SeS/PP/Deg User Manual

Page 52

Advertising
background image

3.14 Metrohm CO

2

Suppressor (MCS)

■■■■■■■■■■■■■■■■■■■■■■

42

■■■■■■■■

930 Compact IC Flex Oven/SeS/PP/Deg (2.930.2560)

ymer membrane. At the same time, the vacuum pump generates a vac-
uum and draws in the air from the outside. The pressure and concentra-
tion difference this creates in the degassing cell relative to inside the capil-
lary causes the CO

2

to be diffused out of the eluent stream. The ambient

air is drawn in by a CO

2

adsorption cartridge that filters out the CO

2

.

3.14.2

Connecting the MCS

The MCS is connected between the suppressor and the detector.

Connecting the MCS

Figure 15

Connecting the MCS

1

Air aspiration capillary
For drawing in air with low CO

2

content (via

the CO

2

adsorption cartridge).

2

Pressure screw, short (6.2744.070)
Installed on the air aspiration capillary.

3

Luer coupling (6.2744.120)
Mounted on the air aspiration capillary with
a pressure screw (6.2744.070).

1 Removing the threaded stoppers

Remove and keep the two threaded stoppers (6.2744.220) from the
input and output of the MCS.

2 Connection from the suppressor

Use a long pressure screw (6.2744.090) to connect the capillary
labeled out to the input of the of the MCS (labeled In).