Advertising
background image

3.3 Removing transport locking screws

■■■■■■■■■■■■■■■■■■■■■■

16

■■■■■■■■

930 Compact IC Flex Oven/SeS/PP/Deg (2.930.2560)

3.3

Removing transport locking screws

To avoid damage to the drives for the high-pressure pump and the vac-
uum pump during transport, the pumps are secured with transport lock-
ing screws. These are located at the rear of the instrument and labeled
with Transport security screws.

Remove these transport locking screws before the initial start-up.

Accessories

For this step you need:

Hex key 4 mm (6.2621.030)

Remove the transport locking screws

Figure 6

Removing the transport locking screws

1

Transport locking screws
For the vacuum pump. Only present if the
instrument has a degasser.

2

Transport locking screws
For the high-pressure pump.

1 Remove all of the transport locking screws with the hex key.

Store the transport locking screws in a safe place. Reinsert the transport
locking screws each time you transport the instrument a significant dis-
tance.