Metrohm 930 Compact IC Flex Oven/SeS/PP/Deg User Manual
Page 32
Advertising
3.6 Connecting the eluent bottle
■■■■■■■■■■■■■■■■■■■■■■
22
■■■■■■■■
930 Compact IC Flex Oven/SeS/PP/Deg (2.930.2560)
4 Mounting the adsorber tube
NOTE
Depending on the eluent used, the adsorber tube (6.1609.000)
must be filled differently:
■
For alkaline eluents or eluents with a low buffer capacity: first a
little cotton, then CO
2
adsorber material.
■
For all other eluents: only cotton.
■
Remove the plastic cover from the large opening of the adsorber
tube. Fill the adsorber tube and close it again using the plastic
cover.
■
Insert the adsorber tube into the bottle cap's large opening. Fas-
ten it to the bottle cap using the SGJ clip (6.2023.020).
Advertising