3 Installation

930 Compact IC Flex Oven/SeS/PP/Deg (2.930.2560)

■■■■■■■■

33

3.12.2.1

Installing the eluent path

The eluent path is connected with the capillaries in and out.

1 Connect the capillary labeled with in with the output of the separa-

tion column using a short pressure screw (6.2744.070).

2 Use a long pressure screw (6.2744.090) to connect the capillary

labeled with out to the input of the MCS (see Chapter 3.14.2, page
42)
.

3.12.2.2

Connecting the regeneration solution

The regeneration solution is connected to the regenerant capillary.

Connecting the regeneration solution to the peristaltic pump
For this step you need the following accessories:

Accessory kit: Flex/Vario: ChS (6.5000.030)

Pump tubing (6.1826.420)

Tubing olive with filter and lock (6.2744.180)

Tubing olive (6.2744.034)

Tubing cartridge of the peristaltic pump