3 Preparing the MME pro for use

6.1246.120 Multi-Mode-Electrode pro

■■■■■■■■

11

Place the MME pro into the electrode stand.
The capillary retaining screw is pointing upwards.

2 Loosening the capillary retaining screw

Remove the transport locking stopper.

Loosen the capillary retaining screw either by hand or with the
wrench.
The capillary retaining screw must sit loosely in the thread.

3 Inserting the capillary

Caution

Only hold the capillary by its sides!

Do not exert pressure on the tip of the capillary!

Each capillary is individually packed in a transparent plastic wrapper.

Cut open the plastic wrapper with the scissors at the marking and
unwrap the capillary.

Grab the capillary at its sides, insert it through the capillary retain-
ing screw into the electrode body.

Gently press it through the silicone seal.

4 Fastening the capillary retaining screw

First, fasten the capillary retaining screw by hand until you feel a
slight resistance.

Then slide the wrench from the top over the capillary retaining
screw.
The knobs on the wrench fit exactly into the grooves oft the capil-
lary retaining screw thus holding the wrench in place.

Turn the handle by pushing it clockwise with one finger until the
wrench falls off.
Now the capillary retaining screw is correctly fastened.

Note

Do not apply the wrench a second time.