5.1 Small maintenance

■■■■■■■■■■■■■■■■■■■■■■

26

■■■■■■■■

6.1246.120 Multi-Mode-Electrode pro

Note

Do not apply the wrench a second time.

See also the video Replace capillary in the multimedia guide Electrodes in
Voltammetry
.

5.1.4

Replacing the needle

Caution

In this procedure mercury is used. Always carry out the following steps
in a dry drip pan.

Replacing the needle

To replace the needle you need the MME pro and the following accesso-
ries:

Drip pan (6.2711.030)

Electrode stand (6.2615.130)

New needle (6.1247.020)

Mounting ring (6.1247.210)

Wrench (6.1247.220)

Scissors