Figure 7, Mme pro – replacing the capillary – Metrohm Multi-Mode Electrode pro User Manual

Page 32

Advertising
background image

5.1 Small maintenance

■■■■■■■■■■■■■■■■■■■■■■

24

■■■■■■■■

6.1246.120 Multi-Mode-Electrode pro

Electrode stand (6.2615.130)

Wrench (6.1247.220)

Silanized capillary (6.1226.050) or non-silanized capillary (6.1226.030)

Scissors

1

2

3

4

Figure 7

MME pro – Replacing the capillary

1 Loosening the needle holder

Place the electrode stand into the drip pan.

Place the MME pro into the electrode stand.

Loosen the needle holder using the wrench turning it counter-
clockwise 1–2 rotations.

The black sealing ring must remain visible!