8 filling in mercury, Filling in mercury – Metrohm Multi-Mode Electrode pro User Manual

Page 47

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Maintenance

6.1246.120 Multi-Mode-Electrode pro

■■■■■■■■

39

4 Fastening the needle holder

Use the wrench to tighten the needle holder by turning it clock-
wise until the black sealing ring becomes visible in the electrode
body.

See also the video Replace the needle in the multimedia guide Electrodes
in Voltammetry
.

5.2.8

Filling in mercury

Caution

In this procedure mercury is used. Always carry out the following steps
in a dry drip pan.

Filling in mercury

To carry out this procedure you need the MME pro and the following
accessories:

Drip pan (6.2744.030)

Electrode stand (6.2615.130)

Wrench (6.1247.220)

Syringe (6.2816.020) with needle (6.2816.030)

Mercury

Mercury drop catcher (6.2406.000)