Figure 9, Mme pro – removing the capillary – Metrohm Multi-Mode Electrode pro User Manual

Page 40

Advertising
background image

5.2 Large maintenance

■■■■■■■■■■■■■■■■■■■■■■

32

■■■■■■■■

6.1246.120 Multi-Mode-Electrode pro

1

2

3

4

Figure 9

MME pro – Removing the capillary

1 Placing the MME pro into the electrode stand

Place the electrode stand into the drip pan.

Place the MME pro into the electrode stand.
The capillary retaining screw is pointing upwards.

2 Removing the capillary retaining screw

Loosen the capillary retaining screw and remove it.

3 Removing the capillary

Pull capillary and the seal out of the MME pro.

4 Removing the seal

Pull the capillary out of the seal.

The seal can be reused if it is undamaged.