Metrohm 940 Professional IC Vario ONE/SeS/PP/Prep 2 User Manual

Page 114

Advertising
background image

6.1

■■■■■■■■■■■■■■■■■■■■■■

104

■■■■■■■■

940 Professional IC Vario ONE/SeS/PP/Prep 2 (2.940.1520)

Problem

Cause

Remedy
Note: 6.2821.180 pump tubing connection
with filter must be used.

Guard column – Blocked.

Replace the guard column (see Chapter 3.22,
page 56)
.

Separation column –
Blocked.

Regenerate the separation column (see
Chapter 5.17.4, page 101)
.

Replace the separation column (see "Con-
necting the separation column", page 60)
.

Note: Samples should always be microfiltered .

Injection valve – Valve
blocked.

Have the valve cleaned (by a Metrohm service
technician).

The retention times
in the chromato-
gram have changed
unexpectedly.

Separation column –
Diminished separating effi-
ciency.

Regenerate the separation column (see
Chapter 5.17.4, page 101)
.

Replace the separation column (see "Con-
necting the separation column", page 60)
.

The eluent contains gas
bubbles.

Ensure that the connectors for the eluent
degasser are firmly connected .

Deaerate the high-pressure pump (see Fig-
ure , page 55)
.

High-pressure pump –
Defective.

Request Metrohm Service.

Peak areas are
lower than
expected.

Sample – There is a leak in
the sample path.

Find and fix the leak in the sample path.

Sample – The sample path
is blocked.

Find and fix the blockage in the sample path.

Sample – The sample loop
is not (completely) filled.

Adjust the transfer time for the sample (see
"Determining the transfer time", page 99)
.

Sample – There are gas
bubbles in the sample.

Use the sample degasser .

MCS – The CO

2

suppressor

is not connected.

Connect the CO

2

suppressor.

The peristaltic pump
is pumping too little.

Peristaltic pump – Contact
pressure too weak.

Correctly set the contact pressure .

Peristaltic pump – Filter
blocked.

Replace the filter (see "Replacing the filter",
page 97)
.