Metrohm 940 Professional IC Vario ONE/SeS/PP/Prep 2 User Manual
Page 71
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
940 Professional IC Vario ONE/SeS/PP/Prep 2 (2.940.1520)
■■■■■■■■
61
2 Installing the input of the separation column
There are 3 possibilities:
■
Attaching the bottom end of the separation column directly to
the guard column.
or
■
If the guard column is connected to the separation column using
a provided connection capillary, connect the bottom end of the
separation column to the guard column's output capillary using
the provided PEEK pressure screw (6.2744.070).
or
■
If no guard column is used (not recommended), connect the col-
umn input capillary to the input of the separation column using a
short pressure screw (6.2744.070).
3 Rinsing the separation column
■
Place a beaker under the separation column's outlet.
■
Start manual control in MagIC Net and select the high-pressure
pump: Manual
▶ Manual control ▶ Pump
– Flow: in accordance with column leaflet
– On
■
Rinse the separation column with eluent for approximately 10
minutes.
■
Use manual control in MagIC Net to stop the high-pressure pump
again: Off.
4 Removing the coupling
■
Remove the coupling (6.2744.040) from the column output capil-
lary.
5 Installing the output of the separation column
■
Fasten the column output capillary to the upper end of the sepa-
ration column using a PEEK pressure screw (6.2744.070).
6 Inserting the separation column
■
Insert the separation column with chip into the column holder
until you hear it snap in place.
The separation column is now detected by MagIC Net.