3.15 Metrohm CO

2

Suppressor (MCS)

■■■■■■■■■■■■■■■■■■■■■■

48

■■■■■■■■

940 Professional IC Vario ONE/SeS/PP/Prep 2 (2.940.1520)

1 Preparing the adsorption cartridge holder

Push the 4 clips into the slots on the adsorption cartridge holder.

2 Connecting adsorption cartridges

Attach the PVC tubing to the tip of the H

2

O adsorption cartridge.

Insert the adapter into the other end of the PVC tubing.

Insert the adapter into the star-shaped cap of the CO

2

adsorption

cartridge.

Attach the capillary connected to the MCS's Air in connection to
the tip of the CO

2

adsorption cartridge.

3 Inserting the adsorption cartridges into the adsorption car-

tridge holder

Clamp the two adsorption cartridges in place using the clips on
the adsorption cartridge holder.

4 Putting the adsorption cartridge holder into the instrument

Place the adsorption cartridge holder with cartridges into the
instrument's detector chamber.