12 metrohm co2 suppressor (mcs), 13 detector, 14 sample degasser – Metrohm 940 Professional IC Vario ONE/SeS/PP/Prep 2 User Manual
Page 121: 15 6-port valve, 12 metrohm co, Suppressor (mcs)
Advertising
■■■■■■■■■■■■■■■■■■■■■■
7 Technical specifications
940 Professional IC Vario ONE/SeS/PP/Prep 2 (2.940.1520)
■■■■■■■■
111
7.12
Metrohm CO
2
Suppressor (MCS)
Material
Fluoropolymer
Resistance to sol-
vents
No restriction (except PFC)
Underpressure
Working range
Microprocessor controlled/stabilized
Time to estab-
lish after start-
ing
< 30 s
Capillary volume
400 µL
Recommended
flow range
0.1 - 1.0 mL/min
7.13
Detector
You can find the technical specifications for the detector in the detector's
manual.
7.14
Sample degasser
Material
Fluoropolymer
Resistance to sol-
vents
No restriction (except PFC)
Time to establish
vacuum
< 60 s
7.15
6-port valve
Actuator switch-
ing time
typ. 100 ms
Maximum operat-
ing pressure
35 MPa (350 bar)
Material
PEEK
Advertising