Figure 3, Thermostat casing – Metrohm 806 Exchange Unit User Manual

Page 23

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

806 Exchange Unit

■■■■■■■■

15

Figure 3

Thermostat casing

Mount the thermostat casing as follows:

1 Loosen the tubing connection to the glass cylinder.

2 Unscrew the light protection.

3 Screw out the glass cylinder with holder.

4 Loosen the screw nipple on the glass cylinder.

5 Roll the O-ring out of the groove and upwards on the glass nozzle.

Do not use any hard items to remove the O-ring, otherwise the edge
of the glass nozzle may chip off!

6 Replace the union nut of the cylinder by the sealing ring of the ther-

mostat casing (threaded side upwards).

7 Slightly grease the O-ring and attach it on the glass nozzle.

8 Fasten the sealing ring with the screw nipple.

9 Insert the glass cylinder with holder into the thermostat casing and

press firmly.

10 Tighten the thermostat casing with the cylinder in the body.

11 Connect the thermostat tubings.