7 Appendix

806 Exchange Unit

■■■■■■■■

31

Table 1

Typical measurement deviation of Metrohm exchange units

Cylinder volume

max. systematic deviation

1 mL

± 3 µL

5 mL

± 15 µL

10 mL

± 20 µL

20 mL

± 30 µL

50 mL

± 50 µL

7.2.2

The ISO/EN/DIN standard 8655-3

The Metrohm exchange units fulfil the requirements of the ISO/EN/DIN
standard 8655-3 Volume measurement instruments with pistons –
Part 3: Piston burets
. Metrohm guarantees that its exchange units are
in compliance with the following limit values at the time of shipment:

Table 2

Permissible limit values as per ISO/EN/DIN 8655-3

Cylinder
volume

max. systematic

measurement deviation

max. permissible

measurement deviation

1 mL

± 0.6 %

± 6 µL

± 0.1 %

± 1 µL

5 mL

± 0.3 %

± 15 µL

± 0.1 %

± 5 µL

10 mL

± 0.2 %

± 20 µL

± 0.07 %

± 7 µL

20 mL

± 0.2 %

± 40 µL

± 0.07 %

± 14 µL

50 mL

± 0.2 %

± 100 µL

± 0.05 %

± 25 µL

The Metrohm agents worldwide offer the possibility of on-site exchange
unit inspections and certifications with respect to accuracy. We recom-
mend that an accuracy inspection be performed when the dosing cylin-
ders and dosing pistons of an exchange unit are replaced.