4 flat stopcock blocked, Flat stopcock blocked – Metrohm 806 Exchange Unit User Manual

Page 32

Advertising
background image

5.1 Care and upkeep

■■■■■■■■■■■■■■■■■■■■■■

24

■■■■■■■■

806 Exchange Unit

7 Couple the piston rod with the push rod of the drive. The clamp of

the push rod must tightly fit into the opening of the piston rod, see
picture.

8 Carefully push the light protection downwards and tighten in the

thread of the body. Thereby the dosing piston is pressed into the cyl-
inder.

9 Dose manually and move the piston upwards to the limit stop.

5.1.4

Flat stopcock blocked

1 Check whether the stopcock is in dosing position. The switching lever

must be directed to the right.

2 Loosen the flat stopcock out of the holder. Rotate the plate holder to

the left and slightly lift the clip on the side of the holder with a finger
nail or a sharp object. Then the flat stopcock can easily be removed
from the holder.

3 Put the stopcock into a solvent.

with aqueous solutions – hot water

with non-aqueous solutions – suitable solvent (see chapter 8)

with KF reagent - methanol, then water, then methanol.

4 Eventually clean the stopcock with a suitable solvent (see above) in

an ultrasonic bath and let it dry afterwards.

5 When the stopcock can be moved again, it can be inserted back into

the exchange unit, see above.