Advertising
▪▪▪▪▪▪▪
23
3.3
Weight
System Type
Pounds
Kilograms
Basic System:
93 lbs
41.9
With Vortec Cooler:
106 lbs
47.7
With Purge Control and Air Conditioning:
135 lbs
60.8
3.4
Operating Temperature Range
System Type
Fahrenheit
Centigrade
Basic System:
32-95
0-35
With Vortec Cooler:
32-104
0-40
With Purge Control and Air Conditioning:
32-125
0-51
With Air Conditioning
32-125
0-51
3.5
Operating Humidity Range
System Type
% Relative Humidity
All Systems:
10-90 % non-condensing
3.6
Power Requirements
•
AC Mains Power: 100/240 VAC Selectable, 50/60 Hz, 2A / 1.5A
•
Power Consumption: 100W Typical
•
Heater adds up to 150W
•
Fans add 20W
•
With Air Conditioning (optional) 5A / 2.7A