3 reference standardization, transmission probes, Reference standardization, transmission probes – Metrohm NIRS XDS Process Analyzer – DirectLight/NonContact User Manual

Page 89

Advertising
background image

▪▪▪▪▪▪▪

87

After the linearization is successfully sent to the
instrument, this message confirms the transfer.

Click “OK” to proceed.

11.4.3

Reference Standardization, Transmission Probes

Reference Standardization with Transmission Probes uses an innovative means to adjust each channel
to a repeatable, transferable level of spectroscopic response. The fiber ends are placed into a fixture,
the response is measured, and the signal is “normalized” using a spectral correction.

This shows the fixture. When Reference
Standardizing, this fixture holds the two probes in
alignment, at a pre-determined distance.

The optical response of the air gap is measured
and stored. A correction is applied, to bring the
channel to “0 Absorbance” in transmission,
corresponding to the response of an air reference.

This fixture is also used in Instrument Calibration.

Follow these instructions to perform Reference Standardization:

1.

Insert each probe into the fixture as

shown.

2.

Tighten the thumbscrew gently on the

probe.

Advertising