3 Installation

IC equipment: Dialysis

■■■■■■■■

13

Tighten a PTFE capillary (6.1803.040) to the inlet of the pump
tubing with orange/yellow stoppers (6.1826.320) using a pressure
screw (6.2744.070).
Immerse the other end of the PTFE capillary in the vessel contain-
ing the acceptor solution (6.1808.070 with 6.1602.150) and fas-
ten with a PVDF pressure screw (6.2744.000).

Tighten another PTFE capillary (6.1803.040) to the outlet of the
pump tubing with orange/yellow stoppers using a pressure screw
(6.2744.070).
Tighten the other end of the PTFE capillary to the acceptor inlet of
the dialysis cell (3-5) using a pressure screw (6.2744.000).

6 Connecting the outlet of the acceptor solution

Tighten the PEEK capillary, 0.5 mm ID / 40 cm (6.1831.050), to
the acceptor outlet of the dialysis cell (3-6) using a pressure screw
(6.2744.000).
Tighten the other end of the PEEK capillary to the sample inlet of
the injection valve (Port 1) using a pressure screw (6.2744.010).

Tighten a PTFE capillary, 0.5 mm ID / 1 m (6.1803.040), to the
sample outlet of the injection valve (Port 2) using a pressure screw
(6.2744.010).
Tighten the other end of the PTFE capillary to the inlet of the sec-
ond pump tubing with orange/yellow stoppers (6.1826.320) using
a pressure screw (6.2744.070).

Tighten another PTFE capillary, 0.5 mm ID / 1 mm (6.1803.040),
to the outlet of the second pump tubing with orange/yellow stop-
pers (6.1826.320) using a pressure screw (6.2744.070).
Tighten the other end of the PTFE capillary to the waste container.