4.1 Operation

■■■■■■■■■■■■■■■■■■■■■■

18

■■■■■■■■

IC equipment: Dialysis

2 Measuring the standard solution with dialysis

Set the dialysis time in the software to 5 minutes.

Immerse the sample aspiration capillary in the standard solution
used in step 1.

Start a determination in the software and wait until the chroma-
togram has been evaluated.

3 Additional measurements

Increase the dialysis time in the software by 1 minute each time
and start the determination.

Repeat the measurements until the measured values are stable.

4 Determining the optimum dialysis time

The recovery rates can be calculated from the result without dialy-
sis and the results with dialysis. You can determine the optimum
dialysis time by recording the recovery rate in relation to the dialy-
sis time.

4.1.2

Recommended procedure for the dialysis

For sample determination with dialysis, it is best to proceed according to
the following sequence:

Carrying out a dialysis

1 Preparing the acceptor solution

To avoid disruptions caused by air bubbles in the acceptor flow,
degas the ultrapure water used as acceptor solution for at least
10 minutes.

2 Preparing the sample

To avoid blockage of the sample channel, always centrifuge sam-
ples with a high content of suspended or solid particles in a tab-
letop centrifuge for 5 minutes at 10,000 rpm.

3 Putting the IC system into operation

Switch on all required devices.

Start the software and load the required IC system.

Condition the IC system.