4.2 Maintenance

■■■■■■■■■■■■■■■■■■■■■■

20

■■■■■■■■

IC equipment: Dialysis

4.2

Maintenance

The dialysis membrane may need to be replaced:

if the signal intensity during dialysis decreases.

if the dialysis membrane dries out.

if the dialysis membrane becomes damaged as a result of depositions
or bacterial growth.

if the sample channel is blocked (i.e., the sample can no longer be
pumped through the dialysis cell).

Replacing the dialysis membrane

1 Disassembling the dialysis cell

Unscrew the four capillaries from the dialysis cell and remove the
cell from the cell holder.

Remove the five screws completely with the hex key (6.2621.070)
and disassemble the dialysis cell.

Remove the used dialysis membrane.

2 Preparing the dialysis membrane

Carry out steps 2 to 8 in Chapter 3.1 Preparing the dialysis cell on
page 6
.

3 Connecting the dialysis cell

Tighten the sample inlet capillary to the dialysis cell inlet for the
sample (3-4) using a PVDF pressure screw (6.2744.000).

Tighten the sample outlet capillary to the dialysis cell outlet for
the sample (3-3) using a PVDF pressure screw (6.2744.000).

Tighten the acceptor inlet capillary to the dialysis cell inlet for the
acceptor solution (3-5) using a PVDF pressure screw (6.2744.000).

Tighten the acceptor outlet capillary to the dialysis cell outlet for
the acceptor solution (3-6) using a PVDF pressure screw
(6.2744.000).

4 Conditioning the dialysis system

(see Chapter 3.4, page 15).