6 Appendix

859 Titrotherm

■■■■■■■■

23

Warning

The temperature sensor is very sensitive and can be considerably dam-
aged by contact with fingers or objects.

Warning

The PTFE protective sleeve of the 6.9011.020 cannot and must not be
dismantled.

The sensor must always be stored in a dry place. After each sample series,
rinse the sensor with water or other cleaning solutions.

6.2

Maximum dosing and filling rate

The maximum dosing rate and maximum filling rate for the dosing unit
depend on the cylinder volume:

Cylinder volume

Maximum rate

2 mL

6.67 mL/min

5 mL

16.67 mL/min

10 mL

33.33 mL/min

20 mL

66.67 mL/min

50 mL

166.00 mL/min

Independent of the cylinder volume, values ranging from 0.01 to 166.00
mL/min can always be entered. When the function is carried out the rate
will be, if necessary, decreased automatically to the highest possible value.