2 maximum dosing and filling rate, Maximum dosing and filling rate – Metrohm 859 Titrotherm User Manual
Page 31
■■■■■■■■■■■■■■■■■■■■■■
6 Appendix
859 Titrotherm
■■■■■■■■
23
Warning
The temperature sensor is very sensitive and can be considerably dam-
aged by contact with fingers or objects.
Warning
The PTFE protective sleeve of the 6.9011.020 cannot and must not be
dismantled.
The sensor must always be stored in a dry place. After each sample series,
rinse the sensor with water or other cleaning solutions.
6.2
Maximum dosing and filling rate
The maximum dosing rate and maximum filling rate for the dosing unit
depend on the cylinder volume:
Cylinder volume
Maximum rate
2 mL
6.67 mL/min
5 mL
16.67 mL/min
10 mL
33.33 mL/min
20 mL
66.67 mL/min
50 mL
166.00 mL/min
Independent of the cylinder volume, values ranging from 0.01 to 166.00
mL/min can always be entered. When the function is carried out the rate
will be, if necessary, decreased automatically to the highest possible value.