2 supplemental characteristics, Maximum input power 350va – B&K 1787B User Manual

Page 13

Advertising
background image

13

Voltage

<0.05% + 10mV

Programming

Accuracy, 12months,
(@ 25 °C ± 5 °C)
±(% of output+offset)

Current

<0.2% +10mA

Voltage

<0.05%+15mV(<20V),

<0.05%+120mV(=

20V)

Readback Accuracy
12months,

(25 °C ± 5 °C)
±(% of output+offset)

Current

<0.1% +15mA

Voltage

=

1mVrms/3mVp-p

Ripple
(20Hz ~20MHz)

Current

5mArms

Voltage

<0.02%+5mV

Temperature
Coefficient,

(0 °C ~ 40 °C)
±(% of output+offset)

Current

<0.1% +5mA

Voltage

<0.02%+15mV(<20V),

<0.02%+120mV(=

20V)

Readback
Yemperature,

Coefficient,
±(% of output+offset)

Current

<0.1% +15mA


2.2 Supplemental Characteristics

State Storage Memory
Sixteen (16) user-configurable stored states

Recommended Calibration Interval
1 year

AC Input Ratings (selectable via switch on the rear panel)
Option OP1: 220VAC ± 10%, 47 to 63 Hz
Option OP2: 110 VAC ± 10%, 47 to 63 Hz

Maximum input power
350VA

Cooling
Fan cooled


Operating Temperature
32 to 104 °F (0 to 40 °C) for full rated output

Storage Temperature
-68 to 158 °F (-20 to 70 °C) for storage environment.